DCM267N
00 (04/11/2008) 4 Velleman
®
Avoid body contact with ground potential (e.g. metallic terminals, output
sockets, lead clamp…) while measuring. Make sure to be electrically
insulated from ground during measurement.
Calibration and repair must be performed by a qualified technician. Refer to
your local dealer.
4. Features
LCD display with automatic polarity indication
data-hold function
DCV (max. 600V) and ACV (max. 600V) measurements
ACA (max. 1000A) current measurements
max. Ø for conductor: 30mm
resistance measurements: max. 40Mohm
frequency measurements: max. 10MHz
temperature measurements: -50°C to 1,000°C
diode test, continuity test and backlight
K-type thermocouple measuring range: from -20°C to 1,000°C
optional probes (not incl.):
SONDE890, SONDE8264
duty cycle measurements : 0.5 to 99% @ pulse width of 100µs to 100ms
capacitance measurement: max. 100µF
safety: indoor use only (overvoltage Cat. II 600V)
5. Overview
Refer to the illustration on page 2 of this manual
1 current clamp 6 Hz - % select button
2 clamp trigger 7 range select button
3 rotary function switch 8 LCD display
4 data hold & backlight button 9 COM jack
5 mode select button 10
V °C/°F jack
6. Operation
GENERAL
When the range of the measured signal is unknown, start measurement
in auto-range mode, and then select range according to indicated value.
The display shows “OL” when the present range limit is exceeded. Select
a higher range to continue measurement.
Do not measure voltages > 600 V or current higher than 1000A. Never
turn the rotary function switch [3] during measurement.
The device will go in sleep mode when no activity is detected for ±30
min.
For all measurements: pressing the data hold button [4] freezes the
currently measured value until the data hold button [4] is pressed again.
The hold-mode is indicated by the word “HOLD” in the display.
To switch on the backlight, press and hold the backlight button [4] for
about one second. To switch the backlight off, press and hold the
backlight button [4] again for ±1s. Note that the meter goes in HOLD
mode when switching the backlight on or off. Press the data hold button
[4] again to continue normal measurement.
The meter starts in auto ranging mode. To select the range manually,
press the range button [7] (not for AC current, diode and continuity
measurements). To return to auto ranging, press and hold the range
button [7] for ±2 seconds.