DVM68
25.02.2011
11
Overrange indication
" OL " displayed automatically
Low Battery indication
" " displayed
Operating temperature
0°C to 40°C
Storage temperature
-
10°C to 50°C
Dimensions
91 x 189 x 31.5 mm
Weight
310 g (incl. battery)
9.1 DC Voltage
Range
Resolution
Accuracy
326mV
0.1mV
± 0.5% of rdg ±
2 digits
3.26V
1mV
32.6V
10mV
± 0.3% of rdg ±
2 digits
326V
0.1V
1000V
1V
±
0.5% of rdg
±
2 digits
Input impedance : 10M
, more than 100M
at 326mV range
9.2 AC Voltage
Range
Resolution
Accuracy
3.26V
1mV
± 0.8% of rdg ±
3 digits
32.6V
10mV
326V
0.1V
750V
1V
Input impedance : 10M
Frequency range : 40 to 1000Hz, 40 to 200Hz at 3.26V range
9.3 DC Current
Range
Resolution
Accuracy
Burden Voltage
326µA
0.1µA
± 1.2% of rdg ± 3 digits
3260µA
1µA
32.6mA
10µA
326mA
0.1mA
10A
10mA
±
2.0% of rdg
±
5 digits
Overload protection :
F 300mA fuse for µA and mA ranges,
F 10A fuse for A range
9.4 AC Current
Range
Resolution
Accuracy
Burden Voltage
326µA
0.1µA
± 1.5% of rdg ± 5 digits
3260µA
1µA
32.6mA
10µA
326mA
0.1mA
10A 10mA
±
3.0% of rdg
±
7 digits
Overload protection :
F 300mA fuse for µA and mA ranges,
F 10A fuse for A range
Frequency range : 40Hz to 1000Hz
9.5 Resistance
Range
Resolution
Accuracy
326
0.1
± 0.8% of rdg ±
3 digits
3.26k
1
32.6k
10
± 0.8% of rdg ±
3 digits
326k
100
3.26M
1k
32.6M
10k
±
1.2% of rdg
±
2 digits
Maximum Open Circuit Voltage : 1.3V
Rev. 01
©Velleman nv
" OL " displayed automatically
2 digits
2 digits
2 digits
3 digits
Burden Voltage
0.5mV / µA
0.5mV / µA
8.0mV / mA
8.0mV / mA
0.02V / A
F 10A fuse for A range
Burden Voltage
0.5mV / µA
0.5mV / µA
8.0mV / mA
8.0mV / mA
0.02V / A
F 10A fuse for A range
3 digits
3 digits
2 digits
Page view 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 54 55

Comments to this Manuals

No comments