1.9. Sensor reconditioning
Over time, the internal sensor may become compromised as a result of pollutants,
chemical vapours or other environmental conditions. This can lead to inaccurate
readings. To recondition the internal sensors, use following procedure:
Heat the datalogger to +80°C (+176°F), humidity < 5% for 36 hours.
Keep the datalogger at+20~+30°C (+70~+90°F), humidity >74% for 48 hours
1.10. Technical Specification
relative humidity (RH) Overall range 0% ~ 100%
relative humidity accuracy 0~20%
20~40%
40~60%
60~80%
80~100%
±5.0%
±3.5%
±3.0%
±3.5%
±5.0%
temperature Overall range -40°C ~ +70°C (-40°F ~ +158°F)
temperature accuracy -40~-10°C (-10~+40°F)
-10~+40°C (+14~+104°F)
+40~+70°C (+104~+158°F)
±2°C (±3.6°F)
±1°C (±1.8°F)
±2°C (±3.6°F)
dew point temperature Overall range -40°C ~ +70°C (-40°F ~ +158°F)
dew point accuracy 25°C, 40~100% RH ±2.0°C (±4.0°F)
logging rate Selectable interval, 2sec ~ 24 hours
operating temperature -35°C ~ +80 °C (-31°F ~ +176°F)
battery type 3.6V Lithium AA)
dimensions 101 x 25 x 23 mm
weight 172 gr. (6 oz) (battery)
Use this device with original accessories only. Velleman nv cannot be held
responsible in the event of damage or injury resulted from (incorrect) use of
this device.
For more info concerning this product, please visit our website
www.velleman.eu.